You routinely purify your samples on an open glass column and have to cope with a limited resolution and a long separation time. As a first step towards automation, we offer two small systems: the “Easy Synthesis” and the “Easy Extract” systems allow you to control the flow rate and the eluent composition.
Improved separation
The use of two pump modules enables the formation of a binary gradient. The solvents enter the pump heads via the aspiration tubing (input) (solvent A and B). The FEP tubing guides the solvent to the mixing chamber with integrated pressure sensor and from there via the injection system on to the column via the pressure tubing (output). The Pump Manager C-615 regulates the flow rate of the pump modules 1 and 2 and monitors the current pressure. The maximum pressure for Pump Module C-601 is 10 bar (145 psi) and for Pump Module C-605 50 bar (725 psi). In addition, a variety of programs and settings enables the selection of flow rate, pressure, time and gradient of the solvents.
Prepacked flash cartridges filled with silica or functionalized silica (C18, Diol, Amine, Cyano, cation exchange), from 4 g up to 330 g (1600 g for silica and C18).
Empty glass columns designed for sample amounts from 1 g up to more than 300 g and pressures up to 50 bar (725 psi) depending on column size.
Quick, efficient and easy packing of plastic or glass cartridges with the packing of your choice.